Multifunction Calibrator
Save Calibration Reports
45
Save Calibration Reports
Use the Product Calibration Report menu to create Calibration Reports and
export the reports to a USB flash drive. The subsequent sections describe the
reports.
To save a calibration report:
1. Connect a flash drive to the USB port on the front of the Product.
2. Go to Setup Menu>Calibration>Generate Report.
3. Select one of the three reports:
• Save Calibration Shifts
• Save Check Shifts
• Save Raw Data
Once an item is selected, the report is saved to the USB flash drive. The report is
in comma separated value (CSV) format.
4. Open or print the file from the PC.
Calibration Shift Results
The Calibration Shift report shows all adjustments made to stored zero-offset and
gain constants, as a result of any drift detected at the most recent calibration.
This report is available and valid at any time.
The report contains this information:
• Heading that shows the date and time the report was saved, the user report
string, and the installed version of the Product software.
• Modules present (installed or attached) including any attached amplifiers
• Most recent calibration dates and temperatures
• Values of externally-calibrated internal references including the resistance
reference, which is the average of the actual-to-nominal ratios of the 100 Ω,
190 Ω, 1 kΩ, 1.9 kΩ, 10 kΩ, 19 kΩ, 100 kΩ, and 190 kΩ resistors.
• DC voltage output shifts for each range
• AC voltage output shifts for each range. A gain shift for all frequencies is
shown along with shifts at selected frequencies for each range.
• Current output shifts for each range (ac and dc)
• Resistance shifts for each value
• 5725A Amplifier output shifts (if attached)
• Wideband AC Module (Option 5700A-03) output shifts (if installed)