Multifunction Calibrator
Full Verification
55
Table 6. 10 Ω through 19 MΩ Verification
Product
Setting
Ohms
DMM
Range
Ohms
2W/4W
Connection
742A
STD Used
742A STD
Ohms
Value
(Rstd)
742A STD
Ohms DMM
Reading
Product
Ohms
DMM
Reading
Calculated
Value
Product
Displayed
Value
Calculated
Value Minus
Displayed
Value
Limits
Ω
10
20 4 Wire 10 Ω
±250 μ
19 ±475 μ
100
200 4 Wire 100 Ω
±1.1 m
190 ±2.09 m
1 K
2 k 4 Wire 1 kΩ
±7.2 m
1.9 K ±13.7 m
10 K
20 k 4 Wire 10 kΩ
±70 m
19 K ±133 m
100 K
200 k 4 Wire 100 kΩ
±800 m
190 K ±1.9
1 M
20 M 2 Wire 1 MΩ
±14
1.9 M ±32.3
10 M
20 M 2 Wire 10 MΩ
±370
19 M ±893
100 M 200 M 2 Wire 100 MΩ ±11 k
Equation 1:
Reading DMM Ohm STD742A
Reading DMM Ohms Product
RstdR =