EasyManua.ls Logo

Fluke 5730A - Page 69

Fluke 5730A
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multifunction Calibrator
Full Verification
59
15. If the readings in step 14 are not the same then adjust the BALANCE control
until they are the same (see note 4 below).
16. Set the CALIBRATE switch to 100:1+. Adjust the 100:1 CALIBRATE
adjustment until the reading is 0±0.5 μV.
17. Push on the Product.
18. Set the CALIBRATE switch to OPERATE and the MODE switch to the
necessary position.
19. This completes the self-calibration of the 752A.
Notes
1. The 10:1 adjustment must be done before the 100:1
adjustment can be done.
2. Following the “Self-Calibrate a Fluke 752A Divider with a
Fluke 8508A as the Null Detector” procedure above adjusts
the 752A to within 0.09 ppm or less of the nominal ratio for
10:1 and 0.18 ppm or less for 100:1.
3. Be sure the 752A is well stabilized in a constant temperature
environment before calibration starts. The instrument stays
calibrated as long as the temperature remains constant.
4. It can be difficult to adjust the BALANCE control for a
difference of exactly 0
μ
V but it is not necessary to get it to
exactly zero to get good results. Instead, adjust the
BALANCE for as close to zero as is practical but a difference
as large as 1
μ
V is still acceptable. Then adjust the 10:1 or
100:1 CALIBRATION adjustment for a reading that is the
same for the switch in the 10:1+ position as the 10:1- position
but opposite sign. For example, if the BALANCE difference is
1
μ
V, then adjust the appropriate CALIBRATION adjustment
until one switch position is say +0.5
μ
v and the other position
is -0.5
μ
V. These two readings do not have to be the same to
meet the
±
0.5
μ
V requirement given in the procedure. To
determine if the adjustment is close enough, use the
following equation: Take the reading in the 10:1+ position
and add it to the reading in the 10:1- position and then divide
the result by two. For example, if the reading in the 10:1+
position is +0.75
μ
V and in the 10:1- position is -0.25
μ
V
then: (+0.75
μ
V+(-0.25
μ
V))/2 = +0.25
μ
V. This meets the
requirement for an adjustment to within ±0.5
μ
V.

Table of Contents

Other manuals for Fluke 5730A

Related product manuals