63
Table 9. DC Voltage Test at Decade Values Verification
10 V
Std
Polarity
Divider
Setting
Product
Range
Product
Output
DMM
Reading
(Vdmm)
Equation
Calculated
Measured
Voltage
Limits
POS 0.1 220 mV 100 mV
dmm
std
V
100
V
±1.2 μV
NEG 0.1 220 mV -100 mV
dmm
std
V
100
V
±1.2 μV
POS 1 2.2 V 1 V
dmm
std
V
10
V
±4.8 μV
NEG 1 2.2 V -1 V
dmm
std
V
10
V
±4.8 μV
POS 10 11 V 10 V
dmmstd
VV +
±33 μV
NEG 10 11 V -10 V
dmmstd
VV +
±33 μV
POS 10 22 V 10 V
dmmstd
VV +
±35 μV
NEG 10 22 V -10 V
dmmstd
VV +
±35 μV
POS 100 220 V 100 V
()
10VV
dmmstd
+
±450 μV
NEG 100 220 V -100 V
()
10VV
dmmstd
+
±450 μV
POS 100 1100 V 100 V
()
10VV
dmmstd
+
±1.1 mV
NEG 100 1100 V -100 V
()
10VV
dmmstd
+
±1.1 mV
POS 1000 1100 V 1000 V
()
100VV
dmmstd
+
±6.5 mV
NEG 1000 1100 V -1000 V
()
100VV
dmmstd
+
±6.5 mV
Vstd=