Multifunction Calibrator
Full Verification
79
Table 17. AC Current 2 mA to 2 A Verification
Product
Applied
Frequency
A40B
Shunt
A40B AC
Resistance
(Ohms)
5790A
Range
5790A
Reading
(Volts)
Calculated
Current
(Amps)
Limits
2 mA 10 Hz 10 mA 220 mV 610 nA
2 mA 20 Hz 10 mA 220 mV 400 nA
2 mA 40 Hz 10 mA 220 mV 274 nA
2 mA 1 kHz 10 mA 220 mV 274 nA
2 mA 5 kHz 10 mA 220 mV 570 nA
2 mA 10 kHz 10 mA 220 mV 3 μA
20 mA 10 Hz 20 mA 2.2 V 6.1 μA
20 mA 20 Hz 20 mA 2.2 V 4 μA
20 mA 40 Hz 20 mA 2.2 V 2.74 μA
20 mA 1 kHz 20 mA 2.2 V 2.74 μA
20 mA 5 kHz 20 mA 2.2 V 5.1 μA
20 mA 10 kHz 20 mA 2.2 V 28 μA
200 mA 10 Hz 200 mA 2.2 V 61 μA
200 mA 20 Hz 200 mA 2.2 V 40 μA
200 mA 40 Hz 200 mA 2.2 V 26.4 μA
200 mA 1 kHz 200 mA 2.2 V 26.4 μA
200 mA 5 kHz 200 mA 2.2 V 48 μA
200 mA 10 kHz 200 mA 2.2 V 232 μA
2 A 20 Hz 2 A 2.2 V 600 μA
2 A 1 kHz 2 A 2.2 V 600 μA
2 A 5 kHz 2 A 2.2 V 1.02 mA
2 A 10 kHz 2 A 2.2 V 14.2 mA