5790B
Service Manual
1-10
Voltage
Range
Frequency Range
±5 °C of Calibration Temperature
AC/DC Transfer
Mode ± ppm
2 Years
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
2.2 V
10 Hz - 20 Hz 55 60 61 62
20 Hz - 40 Hz 19 26 28 34
40 Hz - 20 kHz 15 20 22 27
20 kHz - 50 kHz 15 21 23 33
50 kHz - 100 kHz 49 50 57
100 kHz - 300 kHz 92 110 160
300 kHz - 500 kHz 220 230 280
500 kHz - 1 MHz 830 890 1200
7 V
10 Hz - 20 Hz 55 60 61 62
20 Hz - 40 Hz 19 27 29 36
40 Hz - 20 kHz 15 20 22 27
20 kHz - 50 kHz 18 23 26 35
50 kHz - 100 kHz 62 64 73
100 kHz - 300 kHz 140 150 180
300 kHz - 500 kHz 360 380 450
500 kHz - 1 MHz 1100 1200 1500
22 V
10 Hz - 20 Hz 55 60 61 62
20 Hz - 40 Hz 19 28 30 37
40 Hz - 20 kHz 15 20 22 27
20 kHz - 50 kHz 18 23 26 35
50 kHz - 100 kHz 62 64 69
100 kHz - 300 kHz 140 150 180
300 kHz - 500 kHz 360 380 450
500 kHz - 1 MHz 1100 1200 1500
70 V
[1]
10 Hz - 20 Hz 55 60 62 63
20 Hz - 40 Hz 19 29 31 39
40 Hz - 20 kHz 15 23 25 34
20 kHz - 50 kHz 22 25 27 39
50 kHz - 100 kHz 64 68 85
100 kHz - 300 kHz 140 150 180
300 kHz - 500 kHz 370 390 490
500 kHz - 1 MHz 1100 1200 1500
220 V
[1]
10 Hz - 20 Hz 55 61 62 64
20 Hz - 40 Hz 19 30 32 40
40 Hz - 20 kHz 15 23 25 34
20 kHz - 50 kHz 24 30 34 49
50 kHz - 100 kHz 66 69 83
100 kHz - 300 kHz 160 170 220
300 kHz - 500 kHz 410 480 680
700 V
10 Hz - 20 Hz
[2]
55 62 63 65
20 Hz - 40 Hz 19 31 33 41
40 Hz - 20 kHz 19 24 25 31
20 kHz - 50 kHz 100 110 140
50 kHz - 100 kHz 390 500 850
1000 V
10 Hz - 20 Hz
[2]
55 62 63 65
20 Hz - 40 Hz 19 31 33 41
40 Hz - 20 kHz 19 24 25 31
20 kHz - 50 kHz
[3]
100 110 140
50 kHz - 100 kHz
[3]
390 500 850
[1] Inputs >100 kHz and with a V*Hz product >2.2E7 are typical.
[2] Typical specification, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
[3] Inputs >30 kHz and >750 V are typical, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.