Introduction and Specifications
Electrical Specifications 1
1-9
Relative AC Voltage
Voltage
Range
Frequency Range
±5 °C of Calibration Temperature
AC/DC Transfer
Mode ± ppm 2
Years
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
2.2 mV
10 Hz - 20 Hz 100 + 1.3 110 + 1.3 110 + 1.3
20 Hz - 40 Hz 54 + 1.3 64 + 1.3 68 + 1.3
40 Hz - 20 kHz 44 + 1.3 57 + 1.3 61 + 1.3
20 kHz - 50 kHz 57 + 2.0 67 + 2.0 110 + 2.0
50 kHz - 100 kHz 79 + 2.5 86 + 2.5 120 + 2.5
100 kHz - 300 kHz 190 + 4.0 230 + 4.0 390 + 4.0
300 kHz - 500 kHz 590 + 6.0 720 + 8.0 1200 + 8.0
500 kHz - 1 MHz 2200 + 6.0 2600 + 8.0 4400 + 8.0
7 mV
10 Hz - 20 Hz 80 + 1.3 83 + 1.3 86 + 1.3
20 Hz - 40 Hz 33 + 1.3 39 + 1.3 45 + 1.3
40 Hz - 20 kHz 29 + 1.3 36 + 1.3 42 + 1.3
20 kHz - 50 kHz 40 + 2.0 44 + 2.0 63 + 2.0
50 kHz - 100 kHz 53 + 2.5 57 + 2.5 72 + 2.5
100 kHz - 300 kHz 110 + 4.0 130 + 4.0 210 + 4.0
300 kHz - 500 kHz 370 + 6.0 450 + 8.0 740 + 8.0
500 kHz - 1 MHz 1600 + 6.0 2000 + 8.0 3400 + 8.0
22 mV
10 Hz - 20 Hz 69 + 1.3 72 + 1.3 75 + 1.3
20 Hz - 40 Hz 34 + 1.3 40 + 1.3 46 + 1.3
40 Hz - 20 kHz 30 + 1.3 36 + 1.3 43 + 1.3
20 kHz - 50 kHz 40 + 2.0 45 + 2.0 64 + 2.0
50 kHz - 100 kHz 53 + 2.5 57 + 2.5 73 + 2.5
100 kHz - 300 kHz 97 + 4.0 110 + 4.0 160 + 4.0
300 kHz - 500 kHz 310 + 6.0 380 + 8.0 610 + 8.0
500 kHz - 1 MHz 1200 + 6.0 1500 + 8.0 2500 + 8.0
70 mV
10 Hz - 20 Hz 60 + 1.5 61 + 1.5 62 + 1.5
20 Hz - 40 Hz 27 + 1.5 30 + 1.5 37 + 1.5
40 Hz - 20 kHz 22 + 1.5 25 + 1.5 34 + 1.5
20 kHz - 50 kHz 34 + 2.0 36 + 2.0 44 + 2.0
50 kHz - 100 kHz 53 + 2.5 54 + 2.5 62 + 2.5
100 kHz - 300 kHz 110 + 4.0 120 + 4.0 170 + 4.0
300 kHz - 500 kHz 270 + 6.0 290 + 8.0 320 + 8.0
500 kHz - 1 MHz 910 + 6.0 970 + 8.0 1200 + 8.0
220 mV
10 Hz - 20 Hz 55 60 + 1.5 61 + 1.5 62 + 1.5
20 Hz - 40 Hz 20 27 + 1.5 29 + 1.5 35 + 1.5
40 Hz - 20 kHz 17 22 + 1.5 24 + 1.5 31 + 1.5
20 kHz - 50 kHz 17 22 + 2.0 24 + 2.0 33 + 2.0
50 kHz - 100 kHz 51 + 2.5 52 + 2.5 59 + 2.5
100 kHz - 300 kHz 100 + 4.0 120 + 4.0 170 + 4.0
300 kHz - 500 kHz 260 + 6.0 290 + 8.0 310 + 8.0
500 kHz - 1 MHz 890 + 6.0 950 + 8.0 1200 + 8.0
700 mV
10 Hz - 20 Hz 55 60 + 1.5 61 + 1.5 62 + 1.5
20 Hz - 40 Hz 20 27 + 1.5 29 + 1.5 34 + 1.5
40 Hz - 20 kHz 15 22 + 1.5 24 + 1.5 31 + 1.5
20 kHz - 50 kHz 15 22 + 2.0 24 + 2.0 33 + 2.0
50 kHz - 100 kHz 51 + 2.5 52 + 2.5 59 + 2.5
100 kHz - 300 kHz 100 + 4.0 120 + 4.0 170 + 4.0
300 kHz - 500 kHz 260 + 6.0 270 + 8.0 310 + 8.0
500 kHz - 1 MHz 890 + 6.0 950 + 8.0 1200 + 8.0