5790B
Service Manual
1-8
Voltage Range Frequency Range
±5 °C of Calibration Temperature
AC/DC Transfer
Mode ± ppm
2 Years
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
22 V
10 Hz - 20 Hz
[2]
200 200 200 200
20 Hz - 40 Hz 63 66 67 70
40 Hz - 20 kHz 21 25 27 31
20 kHz - 50 kHz 44 46 48 53
50 kHz - 100 kHz 80 81 85
100 kHz - 300 kHz 180 190 220
300 kHz - 500 kHz 380 400 470
500 kHz - 1 MHz 1100 1200 1500
70 V
[3]
10 Hz - 20 Hz
[2]
200 200 200 200
20 Hz - 40 Hz 63 67 68 72
40 Hz - 20 kHz 25 30 32 39
20 kHz - 50 kHz 55 56 57 63
50 kHz - 100 kHz 91 94 110
100 kHz - 300 kHz 190 200 220
300 kHz - 500 kHz 400 410 510
500 kHz - 1 MHz 1100 1200 1500
220 V
[3]
10 Hz - 20 Hz 200 200 200 200
20 Hz - 40 Hz 63 67 68 72
40 Hz - 20 kHz 23 29 31 38
20 kHz - 50 kHz 63 67 69 77
50 kHz - 100 kHz 96 98 110
100 kHz - 300 kHz 210 210 260
300 kHz - 500 kHz 440 500 700
700 V
10 Hz - 20 Hz
[4]
200 200 200 200
20 Hz - 40 Hz 92 96 99 110
40 Hz - 20 kHz 36 39 41 47
20 kHz - 50 kHz 120 130 150
50 kHz - 100 kHz 400 500 850
1000 V
10 Hz - 20 Hz
[4]
200 200 200 200
20 Hz - 40 Hz 92 96 99 110
40 Hz - 20 kHz 33 37 38 44
20 kHz - 50 kHz
[5]
120 130 150
50 kHz - 100 kHz
[5]
400 500 850
[1] For 9.5 to 10 Hz, the specifications is ±(1000 ppm of reading + 1.5 μV)
[2] For 9.5 to 10 Hz, the specifications is ±(1000 ppm of reading)
[3] Inputs >100 kHz and with a V*Hz product >2.2E7 are typical.
[4] Typical specification, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
[5] Inputs >30 kHz and >750 V are typical, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
Note: The Product is to be used in controlled environments. For disturbances on the mains power supply of >0.5 Vrms from 10 MHz to
40 MHz, add 5 ppm to the 2.2 V range.