Introduction and Specifications
Electrical Specifications 1
1-7
Voltage Range Frequency Range
±5 °C of Calibration Temperature
AC/DC Transfer
Mode ± ppm
2 Years
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
22 mV
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
290 + 1.3
180 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
810 + 4.0
860 + 6.0
1400 + 6.0
290 + 1.3
190 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
810 + 4.0
890 + 8.0
1700 + 8.0
290 + 1.3
190 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
820 + 4.0
1000 + 8.0
2600 + 8.0
70 mV
10 Hz - 20 Hz
[1]
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
240 + 1.5
120 + 1.5
64 + 1.5
120 + 2.0
260 + 2.5
510 + 4.0
660 + 6.0
1100 + 6.0
240 + 1.5
120 + 1.5
65 + 1.5
130 + 2.0
260 + 2.5
510 + 4.0
670 + 8.0
1100 + 8.0
240 + 1.5
130 + 1.5
69 + 1.5
130 + 2.0
260 + 2.5
530 + 4.0
680 + 8.0
1300 + 8.0
220 mV
10 Hz - 20 Hz
[1]
210 210 + 1.5 210 + 1.5 210 + 1.5
20 Hz - 40 Hz 82 84 + 1.5 85 + 1.5 87 + 1.5
40 Hz - 20 kHz 34 37 + 1.5 38 + 1.5 43 + 1.5
20 kHz - 50 kHz 67 69 + 2.0 69 + 2.0 73 + 2.0
50 kHz - 100 kHz 160 + 2.5 160 + 2.5 160 + 2.5
100 kHz - 300 kHz 240 + 4.0 250 + 4.0 280 + 4.0
300 kHz - 500 kHz 360 + 6.0 380 + 8.0 400 + 8.0
500 kHz - 1 MHz 940 + 6.0 1000 + 8.0 1200 + 8.0
700 mV
10 Hz - 20 Hz
[1]
210 210 + 1.5 210 + 1.5 210 + 1.5
20 Hz - 40 Hz 73 75 + 1.5 76 + 1.5 78 + 1.5
40 Hz - 20 kHz 27 31 + 1.5 33 + 1.5 38 + 1.5
20 kHz - 50 kHz 47 50 + 2.0 51 + 2.0 56 + 2.0
50 kHz - 100 kHz 79 + 2.5 79 + 2.5 84 + 2.5
100 kHz - 300 kHz 160 + 4.0 180 + 4.0 210 + 4.0
300 kHz - 500 kHz 300 + 6.0 300 + 8.0 340 + 8.0
500 kHz - 1 MHz 900 + 6.0 960 + 8.0 1200 + 8.0
2.2 V
10 Hz - 20 Hz
[2]
200 200 200 200
20 Hz - 40 Hz 63 65 66 69
40 Hz - 20 kHz 18 22 24 29
20 kHz - 50 kHz 43 45 46 52
50 kHz - 100 kHz 70 71 76
100 kHz - 300 kHz 150 160 200
300 kHz - 500 kHz 250 260 310
500 kHz - 1 MHz 840 900 1200
7 V
10 Hz - 20 Hz
[2]
200 200 200 200
20 Hz - 40 Hz 63 66 67 70
40 Hz - 20 kHz 18 22 24 29
20 kHz - 50 kHz 44 46 48 53
50 kHz - 100 kHz 80 81 88
100 kHz - 300 kHz 180 190 220
300 kHz - 500 kHz 380 400 470
500 kHz - 1 MHz 1100 1200 1500