5790B
Service Manual
3-52
Verifying the Wideband AC Option
Wideband verification verifies that the 5790B WIDEBAND input (requires Option
5790B/3, 5790B/5, or 5790B/AF) is within tolerance. There are two worksheets
and one test record to facilitate this procedure. You will need 1 copy of Table 3-
15, 8 copies of Table 3-16 (one for each voltage range) and 1 copy of the overall
test record, Table 3-17.
Note
If verifying Option 5790B/3, frequencies above 30 MHz are not
applicable.
Table 3-15. Wideband 1 kHz Gain Verification Worksheet
Range
Nominal @
1 kHz
Measured
by 5790B
5730A Value
(Recorded
in Table 3-8)
5790B Error Lower Limit Upper Limit
7 V 3.2 V
2.2 V 2.0 V
2.2 V 1.0 V
2.2 V 0.64 V
700 mV 320 mV
220 mV 100 mV
70 mV 32 mV
22 mV 10 mV
7 mV 3.2 mV
2.2 mV 1 mV
7 mV 3.2 mV