Calibration Adjustment and Verification
Calibration Verification 3
3-57
Step No.
5790B Range
Test Voltage
Frequency
1 Yr Flatness
Specification
Measured
Flatness
Error (%)
90 Days
Absolute
Error Spec
(%)
Measured
Absolute
Error (%)
98 700 mV 320 mV 23 MHz 0.35 No Spec No Spec
99 700 mV 320 mV 26 MHz 0.35 No Spec No Spec
100 700 mV 320 mV 28 MHz 0.35 No Spec No Spec
101 700 mV 320 mV 30 MHz 0.35 No Spec No Spec
102 700 mV 320 mV 35 MHz 0.35 No Spec No Spec
103 700 mV 320 mV 40 MHz 0.35 No Spec No Spec
104 700 mV 320 mV 45 MHz 0.35 No Spec No Spec
105 700 mV 320 mV 50 MHz 0.35 No Spec No Spec
106 220 mV 100 mV 10 Hz 0.10 0.36
107 220 mV 100 mV 20 Hz 0.10 0.36
108 220 mV 100 mV 50 Hz 0.04 0.36
109 220 mV 100 mV 105 Hz 0.04 0.36
110 220 mV 100 mV 200 Hz 0.04 0.36
112 220 mV 100 mV 2 kHz 0.04 0.36
113 220 mV 100 mV 10 kHz 0.04 0.36
114 220 mV 100 mV 20 kHz 0.04 0.36
115 220 mV 100 mV 50 kHz 0.04 0.36
116 220 mV 100 mV 120 kHz 0.04 0.36
117 220 mV 100 mV 200 kHz 0.04 0.36
118 220 mV 100 mV 500 kHz 0.04 0.36
119 220 mV 100 mV 700 kHz 0.05 No Spec No Spec
120 220 mV 100 mV 1 MHz 0.05 No Spec No Spec
121 220 mV 100 mV 1.2 MHz 0.05 No Spec No Spec
122 220 mV 100 mV 2 MHz 0.05 No Spec No Spec
123 220 mV 100 mV 3 MHz 0.10 No Spec No Spec
124 220 mV 100 mV 4 MHz 0.10 No Spec No Spec
125 220 mV 100 mV 6 MHz 0.10 No Spec No Spec
126 220 mV 100 mV 8 MHz 0.10 No Spec No Spec
127 220 mV 100 mV 9 MHz 0.10 No Spec No Spec
128 220 mV 100 mV 10 MHz 0.10 No Spec No Spec
129 220 mV 100 mV 12 MHz 0.15 No Spec No Spec
130 220 mV 100 mV 15 MHz 0.15 No Spec No Spec