Introduction and Specifications
Electrical Specifications 1
1-11
Absolute AC Non-sine Voltage
Range
[1]
Frequency Range
1-Year, tcal ±5 °C,
±(% of reading + µV)
2.2 mV
10 Hz - 45 Hz 0.1 + 1.3
45 Hz - 1 kHz 0.1+ 1.3
1 kHz - 20 kHz 0.17+ 1.3
20 kHz - 100 kHz 0.5 + 2.5
7 mV
10 kHz - 45 Hz 0.1 + 1.3
45 Hz - 1 kHz 0.1 + 1.3
1 kHz - 20 kHz 0.17 + 1.3
20 kHz - 100 kHz 0.5 + 2.5
22 mV
10 Hz - 45 Hz 0.1 + 1.3
45 Hz - 1 kHz 0.1 + 1.3
1 kHz - 20 kHz 0.17+ 1.3
20 kHz - 100 kHz 0.5 + 2.5
70 mV
10 Hz - 45 Hz 0.1 + 1.5
45 kHz - 1 kHz 0.1 + 1.5
1 kHz - 20 kHz 0.17 + 1.5
20 kHz - 100 kHz 0.5 + 2.5
220 mV
10 Hz - 45 Hz 0.1 + 1.5
45 Hz - 1 kHz 0.1 + 1.5
1 kHz - 20 kHz 0.17 + 1.5
20 kHz - 100 kHz 0.5 + 2.5
700 mV
10 Hz - 45 Hz 0.1 + 1.5
45 Hz - 1 kHz 0.1 + 1.5
1 kHz - 20 kHz 0.17 + 1.5
20 kHz - 100 kHz 0.5 +
2.5
2.2 V
[2]
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1 kHz - 20 kHz 0.17
20 kHz - 100 kHz 0.5
7 V
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1 kHz - 20 kHz 0.17
20 kHz - 100 kHz 0.5
22 V
[2]
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1 kHz - 20 kHz 0.17
20 kHz - 100 kHz 0.5
70 V
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1 kHz - 20 kHz 0.17
20 kHz - 100 kHz 0.5
[1] Specifications apply for non-sinusoidal inputs with crest factor <3.0 and with harmonic content band-limited to <1 MHz.
[2] Crest factor limited to <2.3 for signals greater than 75 % of full scale RMS.