5790B
Service Manual
3-38
Table 3-12. DC Voltage Performance Verification Table
Step
No.
Range Nominal
5790B
Positive
Measurement
5790B
Negative
Measurement
5790B
Average*
Lower Limit Upper Limit
1 200 mV 60 mV 59.99628 mV 60.00372 mV
2 200 mV 200 mV 199.99110 mV 200.00890 mV
3 700 mV 200 mV 199.99230 mV 200.00770 mV
4 700 mV 600 mV 599.97990 mV 600.02010 mV
5 2.2 V 0.6 V 0.59998680 V 0.60001320 V
6 2.2 V 1.0 V 0.99997800 V 1.00002200 V
7 2.2 V 2.0 V 1.99995600 V 2.00004400 V
8 7 V 2 V 1.999956 V 2.000044 V
9 7 V 6 V 5.999868 V 6.000132 V
10 22 V 6 V 5.999850 V 6.000150 V
11 22 V 20 V 19.999500 V 20.000500 V
12 70 V 20 V 19.99940 V 20.00060 V
13 70 V 60 V 59.99820 V 60.00180 V
14 220 V 60 V 59.99826 V 60.00174 V
15 220 V 200 V 199.99420 V 200.00580 V
16 700 V 200 V 199.9922 V 200.0078 V
17 700 V 600 V 599.9766 V 600.0234 V
18 1100 V 600 V 599.9778 V 600.0222 V
19 1100 V 1049 V 1048.9612 V 1049.0388 V
*The average of the absolute measurement of the Product with normal and reversed polarity stimulus applied. 5790B
Average = (Meas1 + |Meas2|)/2
1. Connect the 5730A Output HI/LO to the Product Input 2 HI/LO using the
low thermal leads. Connect the 5790B Guard to the 5730A Guard. Ensure
the Guard to Ground strap on the 5730A is installed. Set the Product for
Input 2, External Guard and auto range. Allow three minutes for thermal
dissipation after leads are connected.