Calibration Adjustment and Verification
Calibration Verification 3
3-41
Step No.
5790B
Range
Test
Voltage (V)
Frequency
90 Day
Absolute
AC Error
Spec.
(±ppm)
1 Yr.
Absolute
AC Error
Spec.
(±ppm)
Measured
Absolute
AC Error
(PPM)
2 Yr AC-DC
Error Spec.
(±ppm)
Measured
AC-DC
Error (PPM)
33 0.07 0.06 10 Hz 265 265 No spec
34 0.07 0.06 20 Hz 145 145 No spec
35 0.07 0.06 40 Hz 89 90 No spec
36 0.07 0.06 1 kHz 89 90 No spec
37 0.07 0.06 20 kHz 89 90 No spec
39 0.07 0.06 50 kHz 153 163 No spec
40 0.07 0.06 100 kHz 302 302 No spec
41 0.07 0.06 300 kHz 577 577 No spec
42 0.07 0.06 500 kHz 760 803 No spec
43 0.07 0.06 1 MHz 1200 1230 No spec
44 0.22 0.06 1 kHz 62 63 34
45 0.22 0.06 1 MHz 1040 1100 No spec
46 0.22 0.2 10 Hz 218 218 210
47 0.22 0.2 20 Hz 92 93 82
48 0.22 0.2 40 Hz 45 46 34
50 0.22 0.2 1 kHz 45 46 34
51 0.22 0.2 20 kHz 45 46 34
52 0.22 0.2 50 kHz 79 79 67
53 0.22 0.2 100 kHz 173 173 No spec
54 0.22 0.2 300 kHz 260 270 No spec
55 0.22 0.2 500 kHz 390 420 No spec
56 0.22 0.2 1 MHz 970 1040 No spec
57 0.7 0.6 10 Hz 213 213 210
58 0.7 0.6 20 Hz 78 79 73
60 0.7 0.6 40 Hz 34 36 27
61 0.7 0.6 1 kHz 34 36 27
62 0.7 0.6 20 kHz 34 36 27
63 0.7 0.6 50 kHz 53 54 47
64 0.7 0.6 100 kHz 83 83 No spec
65 0.7 0.6 300 kHz 167 187 No spec
66 0.7 0.6 500 kHz 310 313 No spec
67 0.7 0.6 1 MHz 910 973 No spec

Table of Contents