Calibration Adjustment and Verification
Calibration Verification 3
3-43
Step No.
5790B
Range
Test
Voltage (V)
Frequency
90 Day
Absolute
AC Error
Spec.
(±ppm)
1 Yr.
Absolute
AC Error
Spec.
(±ppm)
Measured
Absolute
AC Error
(PPM)
2 Yr AC-DC
Error Spec.
(±ppm)
Measured
AC-DC
Error (PPM)
104 7.0 6.0 40 Hz 22 24 18
105 7.0 6.0 1 kHz 22 24 18
106 7.0 6.0 20 kHz 22 24 18
107 7.0 6.0 50 kHz 46 48 44
108 7.0 6.0 100 kHz 80 81 No spec
109 7.0 6.0 300 kHz 180 190 No spec
110 7.0 6.0 500 kHz 380 400 No spec
111 7.0 6.0 1 MHz 1100 1200 No spec
113 22 20.0 10 Hz 200 200 123
114 22 20.0 20 Hz 66 67 123
115 22 20.0 40 Hz 25 27 21
116 22 20.0 1 kHz 25 27 21
117 22 20.0 20 kHz 25 27 21
119 22 20.0 50 kHz 46 48 44
120 22 20.0 100 kHz 80 81 No spec
121 22 20.0 300 kHz 180 190 No spec
122 22 20.0 500 kHz 380 400 No spec
123 22 20.0 1 MHz 1100 1200 No spec
125 70 20.0 500 kHz 400 410 No spec
126 70 20.0 1 MHz 1100 1200 No spec
127 70 60.0 10 Hz 200 200 200
128 70 60.0 20 Hz 67 68 63
129 70 60.0 40 Hz 30 32 25
130 70 60.0 1 kHz 30 32 25
131 70 60.0 20 kHz 30 32 25
132 70 60.0 50 kHz 56 57 55
133 70 60.0 100 kHz 91 94 No spec
134 70 60.0 300 kHz 190 200 No spec
135 220 60 300 kHz 210 210 No spec
136 220 200.0 10 Hz 200 200 200
137 220 200.0 20 Hz 67 68 63
138 220 200.0 40 Hz 29 31 23