6109A/7109A
Operators Manual
38
Security determines which menus and functions require the password.
Regardless of the Security setting, these menus are protected:
Setup > Instrument > More > Control
Setup > Instrument > More > Password
Setup > Instrument > More > Service > System Status > Update Firmware
Setup > Instrument > More > Service > Bath Calibration
Setup > Instrument > More > Service > Input Calibration (-P models)
Setup > Instrument > More > Memory > Clear Settings
Setup > Instrument > More > Memory > Clear Data
Setup > Probe > Reference Probe > Program Probe (-P models)
When Security is ON, these additional menus are protected:
Program > Select > Edit
Program > Select > Options
Program > New
Program > Delete
Program > Reports > Delete
Setup > Operation
Setup > Instrument
Setup > Probe > Reference Probe (-P models)
Setup > Probe > DUT (-P models)
(Setup > Instrument > More > Password) Default
Change the password back to the factory default (1234).
(Setup > Instrument > More) Service
View or change service related parameters.
Product information is shown (Model, Serial Number, Firmware Version,
Bath Calibration date, and Input Calibration date (-P models)).
(Setup > Instrument > More > Service) System Status
View information about the condition of the Product.
The System Status screen shows the results of the latest self-check. A
self-check is performed when the Product is switched on.
(Setup > Instrument > More > Service > System Status) Self Check
Repeat the system self-check.
Subsystems and components are tested. When completed, the System
Status screen shows updated results.
(Setup > Instrument > More > Service > System Status) Diagnostics
View data about the Product operation.
The diagnostics indicators can be helpful when the Product is serviced
and for troubleshooting issues.