Portable Calibration Baths
Commands
49
CALCulate2:AVERage:SDEViation?
(-P models only) Return the moving standard deviation of the DUT sensor
readings.
The response is a temperature value in °C or °F and depends on the temperature
unit setting.
CALCulate2:CONVersion:CATalog? [<function>]
(-P models only) Return the list of supported RTD types or thermocouple types.
<function> is either RTD or TC. If the parameter is omitted, the current function is
assumed.
For RTD, the response is PT_A385,PT_A392,PT_JIS,NI_120,RES.
For TC, the response is C,D,E,G,J,K,L,M,N,R,S,T,U,W,MV.
CALCulate2:CONVersion:TYPE <type>
CALCulate2:CONVersion:TYPE?
(-P models only) Set the DUT sensor type.
<type> is the specific DUT sensor type. The allowed types depend on the current
function (see CALCulate2:CONVersion:CATalog? and SENSE2:FUNCtion).
When the DUT sensor type is changed, the RTD wires setting (see
INPut2:RTD:WIRes) and transmitter loop power setting (see
INPut2:MAMP:LPOWer) revert to their default values.
This command is conditionally password protected.
CALCulate2:DATA?
(-P models only) Return the DUT sensor reading.
If the DUT sensor is configured to produce temperature readings, the response is
in °C or °F and depends on the temperature unit setting.
CALCulate2:SCALe:SPAN <numeric_value>|DEF
CALCulate2:SCALe:SPAN?
(-P models only) Set the mA scaling span.
<n> is the span between the reading at 20 mA and the reading at 4 mA. The
default is 16.
This command is conditionally password protected.