Portable Calibration Baths
Commands
57
SENSe2:DATA?
(-P models only) Return the DUT input measurement as resistance in ohms,
voltage in mV, or current in mA.
SENSe2:FUNction <function>
SENSe2:FUNction?
(-P models only) Select the function for the Input Module DUT inputs.
<function> is either RTD, TC, MA, or NONE.
This command is conditionally password protected.
SOURce:CUTout:CLEAr
Clear cutout condition.
For the cutout to be cleared, the fluid temperature must be less than the cutout
temperature setting.
SOURce:CUTout:EVENt?
Return the most recent cutout event.
The response is a sum of one or more codes. Codes are interpreted as in
Table 8:
Table 8. Cutout Codes
Code Definitions
0 No cutout occurred
1 Fluid temperature exceeded the cutout temperature
2 Tank is empty
16 Heater temperature exceeded the cutout temperature
32 Stir motor failed
SOURce:CUTout:LEVel <numeric_value>
SOURce:CUTout:LEVel?
Set the cutout temperature.
<numeric_value> is the temperature in °C or °F and depends on the temperature
unit setting.
This command is conditionally password protected.
SOURce:CUTout:STATe?
Return the cutout condition.
The response is 1 if the cutout is active, or 0 otherwise.