Commands
61
SOURce:STIR:CALibration <numeric_value>
SOURce:STIR:CALibration?
Set the reference stir power percent.
<numeric_value> is the stir power percent.
Calibration parameters should only be changed by a qualified technician as part
of a calibration procedure.
This command is password protected.
SOURce:STIR:POWer?
Return the stir power percent.
SOURce:STIR:SPEEd <numeric_value>
SOURce:STIR:SPEEd?
Set the stir speed.
<numeric_value> is the stir speed in revolutions per minute (RPM).
Calibration parameters should only be changed by a qualified technician as part
of a calibration procedure.
This command is password protected.
SOURce:STIR:TACHometer?
Return the actual stir speed.
The response is the stir speed in revolutions per minute (RPM). Stir speed is 0 if
control is off.
STATus:OPERation[:EVENt]?
Return and clear the SCPI Operation Status Event register.
(-P models only) The response is 16 (measurement ready) if the latest reading of
the reference probe has not yet been queried (see the command FETCh?).
STATus:OPERation:CONDition?
Return the SCPI Operation Status Condition register.
(-P models only) The response is 16 (Measuring).
STATus:OPERation:ENABle <numeric_value>
STATus:OPERation:ENABle?
Set bits in the SCPI Operation Status Mask register.
<numeric_value> is a decimal number that ranges from 0 to 255. This
determines which bits of the Operation Status Event Register affect the OSB
summary message of the Status Byte Register.