725
Users Manual
24
Table 6. RTD Types Accepted
RTD Type Ice Point (R
0
) Material α Range (°C)
Pt100 (3926) 100 Platinum 0.003926 /°C -200 to 630
Pt100 (385) 100 Platinum 0.00385 /°C -200 to 800
Ni120 (672) 120 Nickel 0.00672 /°C -80 to 260
Pt200 (385) 200 Platinum 0.00385 /°C -200 to 630
Pt500 (385) 500 Platinum 0.00385 /°C -200 to 630
Pt1000 (385) 1000 Platinum 0.00385 /°C -200 to 630
Pt100 (3916) 100 Platinum 0.003916 /°C -200 to 630
The Pt100 commonly used in U.S. industrial applications is Pt100 (3916), α = 0.003916 /°C. (Also designated as JIS
curve.) The IEC standard RTD is the Pt100 (385), α = 0.00385 /°C.