Documenting Process Calibrator
Detailed Specifications
111
+DC Current Source
Range/Mode Resolution
% of Output + Floor
1 Year 2 Year
0.100 to 22.000 mA
1 μA 0.01 % + 3 μA 0.02 % + 3 μA
Temperature Coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
Source mA Compliance Voltage: 18 V maximum
Source mA Open Circuit Voltage: 30 V maximum
+DC Current Simulate (External Loop Power)
Range/Mode Resolution
% of Output + Floor
1 Year 2 Year
0.100 to 22.000 mA (Current Sink)
1 μA 0.02 % + 7 μA 0.04 % + 7 μA
Simulate mA Input Voltage: 15 to 50 V dc, add 300 μA to floor when >25 V is present on the loop
Temperature Coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
Resistance Sourcing
Range Resolution
% of Output + Floor
Allowable Excitation
Current
1 Year 2 Year
10.000 Ω 0.001 Ω 0.01 % + 0.010 Ω 0.015 % + 0.015 Ω
0.1 mA to 10 mA
100.00 Ω
[1]
0.01 Ω 0.01 % + 0.02 Ω 0.015 % + 0.03 Ω
0.1 mA to 10 mA
1.0000 kΩ
[2]
0.1 Ω 0.02 % + 0.0002 kΩ 0.03 % + 0.0003 kΩ
0.01 mA to 1.0 mA
10.000 kΩ 1 Ω 0.02 % + 0.003 kΩ 0.03 % + 0.005 kΩ
0.01 mA to 1.0 mA
Temperature Coefficient: (0.01 % of output +0.02 % of range / °C (<18 °C or >28 °C).
When connected to mains, accuracy is not specified with conducted RF >1V, 8 to 15 MHz.
[1] Add 0.01 Ω when the excitation current is <1 mA.
[2] Add 0.0015 kΩ when the excitation current is <0.1 mA.