EasyManua.ls Logo

Fluke 754 - Temperature, Resistance Temperature Detectors

Fluke 754
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
753/754
Users Manual
116
Temperature, Resistance Temperature Detectors
Temperature, RTDs
Degrees or % of Reading
[1]
Type (α) Range °C
Measure °C
[2]
Source °C
Allowable
Excitation
Current
[3]
1 Year 2 Year
Source
Current
1 Year 2 Year
100 Ω
Pt(385)
-200 to 100
0.07 °C 0.14 °C
1 mA
0.05 °C 0.10 °C
0.1 to 10 mA
100 to 800
0.02 % + 0.05 °C 0.04 % + 0.10 °C 0.0125 % + 0.04 °C 0.025 % + 0.08 °C
200 Ω
Pt(385)
-200 to 100
0.07 °C 0.14 °C
500 μA
0.10 °C 0.20 °C
0.1 to 1 mA
100 to 630
0.02 % + 0.05 °C 0.04 % + 0.10 °C 0.017 % + 0.09 °C 0.034 % + 0.18 °C
500 Ω
Pt(385)
-200 to 100
0.07 °C 0.14 °C
250 μA
0.08 °C 0.16 °C
0.1 to 1 mA
100 to 630
0.02 % + 0.05 °C 0.04 % + 0.10 °C 0.017 % + 0.06 °C 0.034 % + 0.12 °C
1000 Ω
Pt(385)
-200 to 100
0.07 °C 0.14 °C 150 μA 0.06 °C 0.12 °C
0.1 to 1 mA
100 to 630
0.02 % + 0.05 °C 0.04 % + 0.10 °C
0.017 % + 0.05 °C 0.034 % + 0.10 °C
100 Ω
Pt(3916)
-200 to 100
0.07 °C 0.14 °C
1 mA
0.05 °C 0.10 °C
0.1 to 10 mA
100 to 630
0.02 % + 0.05 °C 0.04 % + 0.10 °C 0.0125 % + 0.04 °C 0.025 % + 0.08 °C
100 Ω
Pt(3926)
-200 to 100
0.08 °C 0.16 °C
1 mA
0.05 °C 0.10 °C
0.1 to 10 mA
100 to 630
0.02 % + 0.06 °C 0.04 % + 0.12 °C 0.0125 % + 0.04 °C 0.025 % + 0.08 °C
1.800.868.7495info@Fluke-Direct.ca
Fluke-Direct.ca

Table of Contents

Other manuals for Fluke 754