Table 2. Summary of Source and Measure Functions
Function Measure Source
Volts dc 0 V to ±300 V 0 V to ±15 V (10 mA max)
h Volts ac 0.27 V to 300 V rms, 40 Hz to 500 Hz No sourcing
h Frequency
1 Hz to 50 kHz
0.1 V to 30 V p-p sine wave, or 15 V peak
square wave, 0.1 Hz to 50 kHz sine wave,
0.01 Hz square wave
q Resistance 0 Ω to 10 kΩ 0 Ω to 10 kΩ
dc Current 0 mA to 100 mA 0 to 22 mA sourcing or sinking
q Continuity
Beep and the word Short indicates
continuity
No sourcing
t Thermocouple Types E, N, J, K, T, B, R, S, C, L,U, BP, or XK
t RTD
(2-W, 3-W, 4-W)
100 Ω Platinum (3926)
100 Ω Platinum (385)
120 Ω Nickel (672)
200 Ω Platinum (385)
500 Ω Platinum (385)
1000 Ω Platinum (385)
10 Ω Copper (427)
100 Ω Platinum (3916)
p Pressure
[1]
29 modules ranging from 0 to 1 inch H
2
O
(250 Pa) to 0 to 10,000 psi (69,000 kPa)
s Loop Power 26 V
[1] Use an external hand pump or other pressure source as a pressure stimulus for the source pressure function.