Frequency Sourcing
Range
Specification
2 Year
Sine Wave: 0.1 Hz to 10.99 Hz 0.01 Hz
Square Wave: 0.01 Hz to 10.99 Hz 0.01 Hz
Sine and Square Wave: 11.00 Hz to 109.99 Hz 0.1 Hz
Sine and Square Wave: 110.0 Hz to 1099.9 Hz 0.1 Hz
Sine and Square Wave: 1.100 kHz to 21.999 kHz 0.002 kHz
Sine and Square Wave: 22.000 kHz to 50.000 kHz 0.005 kHz
Waveform Choices: Zero-symmetric sine wave or positive 50 % duty-cycle square wave
Square Wave Amplitude: 0.1 to 15 V p-p
Square Wave Amplitude Accuracy, 0.01 to 1 kHz: 3 % p-p output + 75 mV, 1 kHz to 50 kHz: 10 % p-p output + 75 mV typical.
Sine Wave Amplitude: 0.1 to 30 V p-p
Sine Wave Amplitude Accuracy, 0.1 to 1 kHz: 3 % p-p output + 75 mV, 1 kHz to 50 kHz: 10 % p-p output + 75 mV typical.
Frequency specifications are valid when averaged ≥100 ms