Documenting Process Calibrator
Operation Features
13
Operation Features
Input and Output Jacks
Figure 4 shows the input and output jacks and connectors.
Table 3 explains their use.
Table 3. Input/Output Jacks and Connectors
No. Name Description
HART jack (754 only) Connects the Product to HART devices.
Pressure module
connector
Connects the Product to a pressure module.
TC input/output
Jack to measure or simulate thermocouples. This jack accepts a miniature
polarized thermocouple plug with flat, in-line blades spaced 7.9 mm (0.312 in)
center to center.
,
MEASURE V jacks
Input jacks to measure voltage, frequency, or three- or four-wire RTDs
(Resistance Temperature Detectors).
,
SOURCE mA,
MEASURE mA Ω RTD
jacks
Jacks to source or measure current, measure resistance and RTDs, and supply
loop power.
,
SOURCE V Ω RTD
jacks
Output jacks to source voltage, resistance, frequency, and to simulate RTDs.
Battery Charger jack
Jack for the battery charger/universal power supply (referred to as the battery
charger throughout this manual). Use the battery charger for bench-top
applications where ac line power is available.
USB port (Type 2) Connects the Product to a USB port on a PC.