77 Series III
Service Manual
3-10
Table 3-4. Resistance Test
Step Input Resistance Display Reading
1 short 0.0 to 0.3
Decades of 1:
2 100 99.2 to 100.8 (plus 0 reading)
31k 994 to 1006
4 10 k 9.94 to 10.06 k
5 100 k 99.4 to 100.6 k
61 M .994 to 1.006 M
7 10 M 9.79 to 10.21 M
Decades of 1.9:
2 190 188.7 to 191.3 (plus 0 reading)
3 1.9 k 1889 to 1911
4 19 k 18.89 to 19.11 k
5 190 k 188.9 to 191.1 k
6 1.9 M 1.889 to 1.9111 M
7 19 M 18.61 to 19.39 M
Decades of 2.7:
2 270 268.3 to 271.7 (plus 0 reading)
3 2.7 k 2685 to 2715
4 27 k 26.85 to 27.15 k
5 270 k 268.5 to 271.5 k
6 2.7 M 2.685 to 2.715 M
7 27 M 26.45 to 27.55 M
3-11. DC mA Test
1. Set the output of the DC Current Calibrator to zero, and connect it to
the 300mA and COM input terminals of the UUT.
2. Set the DC Current Calibrator to the output shown in Table 3-5, and
verify that the UUT display reading is within the limits shown.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 77 Series III and is the answer not in the manual?

Fluke 77 Series III Specifications

General IconGeneral
Diode TestYes
Continuity BeeperYes
Battery Type9V
Safety RatingCAT III 600V
DisplayLCD, 4000 counts
Voltage DC1000V
Voltage AC1000V
Current DC10A
Current AC10A
Resistance40 MΩ
Weight340 g

Related product manuals