EasyManua.ls Logo

Fluke 787 - Advanced Measurement Features; TouchHold Functionality; Test Lead Resistance Compensation

Fluke 787
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
787
Users Manual
12
Table 4. Rotary Switch Positions for mA Output
No. Position Default Function Pushbutton Actions
OUTPUT
[ mA
Test leads in SOURCE:
Source 0 % mA
Test leads in
SIMULATE:
Sink 0 % mA
% STEP X or W: Adjusts output up or down to the next 25 % step
COARSE Z or Y: Adjusts output up or down 0.1 mA
FINE X or W: Adjusts output up or down 0.001 mA
OUTPUT
mA
J
Test leads in SOURCE:
Source repeating
0 % -100 %-0 %
slow ramp (E)
Test leads in
SIMULATE:
Sink repeating
0 % -100 %-0 %
slow ramp (E)
BLUE cycles through:
Fast repeating 0 % -100 % - 0 % ramp (P on display)
Repeating 0 % -100 % - 0 % ramp in 25 % steps (N on display)
Slow repeating 0 % -100 % - 0 % ramp (E on display)
1.888.475.5235info@Fluke-Direct.com