ProcessMeter™
Specifications
39
AC Current Measurement
Range
45 Hz to 2 kHz
Resolution Accuracy, ±(% of Reading + Counts)
Typical Burden
Voltage
1.000 A (Note) 0.001 A 1 % + 2 1.5 V/A
Note: 440 mA continuous, 1 A 30 seconds maximum on, 5 minutes off
Specifications are valid from 5 % to 100 % of amplitude range.
AC conversion: true rms
Maximum crest factor: 3
For non-sinusoidal waveforms, add
±
(2 % reading + 2 % f.s.) typical
Overload protection 440 mA, 1000V fast-blow fuse
DC Current Measurement
Range Resolution Accuracy, ±(% of Reading + Counts)
Typical Burden
Voltage
30.000 mA 0.001 mA 0.05 % + 2 14 mV/mA
1.000 A (Note) 0.001 A 0.2 % + 2 1.5 V/A
Note: 440 mA continuous, 1A 30 seconds maximum on, 5 minutes off
Overload protection: 440 mA, 1000V fast-blow fuse