Fluke 805 - External Sensor Connection and Sensitivity Setup

Fluke 805
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vibration Meter
How to Operate
15
Note
Fluke supports, but does not provide, external
sensors.
gqi05.eps
Figure 3. External Sensor Connection
Note
High frequency measurement (Crest Factor+) and
Temperature measurement automatically turn off
when an external sensor is connected to the
Meter.
To connect:
1. Open the connector cover and push the external
sensor into position.
After you connect the external sensor, a pop-up
message opens on the Meter display and then
opens the Enter Sensitivity menu.
gqi14.eps
Note
The sensitivity must be set in mV/g units.
2. Push and to select a character in the menu.
3. Push to input the character into the field.
4. Repeat steps 2 and 3 for additional characters.
5. Push to store the value in the Meter and exit
the menu.
6. Push to start data collection.
The Meter automatically detects when you disconnect the
external sensor and is set to measure with the internal
sensor.