805
Users Manual
34
Clear All Data
The Clear ALL Data option lets you delete all the Machine
Setups and measurements.
To clear memory:
1. Push .
2. Push and to highlight Clear ALL Data.
3. Push .
gqi93.bmp
4. Push to select Yes.
5. Push to delete all data.
Interpret Results
The Meter is a screening tool to identify machinery
problems for more diagnostic tests. The Meter has a
vibration severity scale for Bearing and Overall Vibration
measurements. It can also trend vibration measurements
over time. If a measurement shows high vibration severity,
or if there is an adverse trend in vibration severity over
time, then the machine can have a problem. Fluke
recommends that you consult a vibration specialist for
more tests to find the root cause of these problems.
Severity Scale
Measurements with a machine ID, machine category, and
rotational speeds >600 RPM includes both severity scales.
A good measurement always finds some vibration. There
are four severity levels: good, satisfactory, unsatisfactory,
and unacceptable. A measurement in the good category is
an indication of a healthy machine. Table 6 lists the
different severity scales.