80i-110s Instruction Manual 19
AC Measurement:
Table 3. Performance Test Points (5100B and 5220A)
Currrent Range 0 to 100A
DC Measurement:
AC Measurement:
5100B
SETTINGS
RMS AMPS
MEASURED
LOW LIMIT
OUTPUT
HIGH LIMIT
OUTPUT
0.1A, 60 Hz
0.3A, 400 Hz
0.5A, 2 kHz
0.6A, 4 kHz
1A
3A
5A
6A
92 mV
286 mV
465 mV
559 mV
108 mV
314 mV
535 mV
641 mV
5100B
SETTINGS
DC AMPS
MEASURED
LOW LIMIT
OUTPUT
HIGH LIMIT
OUTPUT
1V
3V
5V
7V
9V
10A
30A
50A
70A
90A
95.5 mV
287.5 mV
439.5 mV
615.5 mV
765.0 mV
104.5 mV
312.5 mV
560.5 mV
784.5 mV
1035.0 mV
5100B
SETTINGS
RMS AMPS
MEASURED
LOW LIMIT
OUTPUT
HIGH LIMIT
OUTPUT
1V, 5 kHz
2V, 5 kHz
3V, 2 kHz
5V, 400 Hz
7V, 60 Hz
10A
20A
30A
50A
70A
92.5 mV
185.5 mV
254.5 mV
439.5 mV
595.0 mV
107.5 mV
214.5 mV
345.5 mV
560.5 mV
805.0 mV