Manual Supplement 810 Getting Started
2 3/10
On page 12, replace the Sensor Setup and the Compatible Sensor
paragraphs with the following:
Sensor Setup
The Tester includes a triaxial Sensor with TEDS technology (Transducer Electronic Data Sheets). With this
technology, your Tester can identify and automatically read the Sensor configuration. This technology provides:
‚ Improved results from detailed calibration information
‚ Reduced configuration time without manual data entry
‚ Sensor calibration tracking with last calibration date stored electronically
Compatible Sensors
It is highly recommended to use a Fluke Triaxial Sensor with the Tester. Using a Sensor other than a Fluke
Triaxial Sensor will result in misleading diagnoses. The Tester is compatible with single axis Sensors.
On page 13, add the following bullet under the Caution:
‚ Always disconnect the Sensor cable from the Tester when not in use.
On page 14, delete steps 2, 3 and 4 and add the following Warning:
V Warning
To avoid injury when attaching reflective tape to the machine, stop the rotating machine. Attach a piece
of reflective tape onto the shaft or other rotating part of the machine. Restart the machine and wait until
it reaches its normal operating conditions.