87V MAX
Users Manual
44
DC Current
Function Range Resolution Burden Voltage Accuracy
μA dc
600.0 μA 0.1 μA 100 μV/μA ±(0.2 % + 4)
6000 μA 1 μA 100 μV/μA ±(0.2 % + 2)
mA dc
60.00 mA 0.01 mA 1.8 mV/mA
±(0.2 % + 4)
400.0 mA
[1]
0.1 mA 1.8 mV/mA
±(0.2 % + 2)
A dc
6.000 A 0.001 A 0.03 V/A
±(0.2 % + 4)
10.00 A
[2,3]
0.01 A 0.03 V/A
±(0.2 % + 2)
Capacitance
Range Resolution Accuracy
10.00 nF 0.01 nF
±(1.0 % + 2)
[1]
100.0 nF 0.1 nF
1.000 μF 0.001 μF
±(1.0 % + 2)
10.00 μF 0.01 μF
100.0 μF 0.1 μF
9999 μF 1 μF
[1] With a film capacitor or better, using the rel mode to zero residual.
Diode
Range Resolution Accuracy
2.000 V 0.001 V
±(1.0 % + 1)
1.800.561.8187info@Fluke-Direct.ca