Introduction and Specifications 
  Electrical Specifications  1 
1-13 
 
Uncertainty 
[1]
 
90 days  1 year 
Range 
23 °C ± 5 °C  23 °C ± 5 °C 
Temperature Coefficient/°C  
Outside 18 – 28 °C 
200 μA 
0.02 + 0.005  0.03 + 0.005  0.003 + 0.001 
2 mA  0.015 + 0.005  0.02 + 0.005  0.002 + 0.001 
20 mA  0.03 + 0.02  0.04 + 0.02  0.005 + 0.001 
200 mA  0.02 + 0.005  0.03 + 0.008  0.005 + 0.001 
2 A  0.05 + 0.02  0.08 + 0.02  0.008 + 0.001 
10 A  0.18 + 0.01  0.2 + 0.01  0.008 + 0.001 
Notes: 
[1]   Uncertainty given as ± (% of reading + % of range) 
AC Current 
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs 
from 1 % to 5 % of range, add an additional error of 0.1 % of range. 
Input Protection ................................................ Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses 
Measurement Method ....................................... AC-coupled True RMS 
Shunt Resistance.............................................. 0.01 Ω for 2 A and 10 A ranges 
1 Ω for 20 mA and 200 mA  
AC Filter Bandwidth ......................................... 20 Hz – 100 kHz 
Maximum Crest Factor ..................................... 3:1 at Full Scale 
Additional Crest Factor Errors (<100 Hz)  ...... Crest Factor 1-2, 0.05 % of full scale 
Crest Factor 2-3, 0.2 % of full scale  
Only applies to non-sinusoid signals 
Input Characteristics 
Resolution 
Range 
Full-Scale 
(5-1/2 Digits) 
Slow Medium  Fast 
Burden Voltage 
 
20 mA  19.9999 mA 
0.1 μA 1 μA 1 μA 
<0.05 V 
200 mA  199.999 mA 
1 μA 10 μA 10 μA 
<0.5 V 
2 A  1.99999 A 
10 μA 100 μA 100 μA 
<0.1 V 
10 A  10.0000 A 
100 μA 
1 mA  1 mA 
<0.5 V 
 
Uncertainty 
[1]
 
90 days  1 year 
Range Frequency 
23 °C ± 5 °C  23 °C ± 5 °C 
Temperature 
Coefficient/°C 
Outside 18 – 28 °C
20 mA  20 Hz - 45Hz  1 + 0.05  1.25 + 0.06  0.015 + 0.005 
  45 Hz - 2 kHz  0.25 + 0.05  0.3 + 0.06  0.015 + 0.005 
200 mA  20 Hz - 45Hz  0.8 + 0.05  1 + 0. 06  0.015 + 0.005 
  45 Hz - 2 kHz  0.25 + 0.05  0.3 + 0.06  0.015 + 0.005 
2 A  20 Hz - 45Hz  1 + 0.05  1.25 + 0.06  0.015 + 0.005 
  45 Hz - 2 kHz  0.25 + 0.05  0.3 + 0.06  0.015 + 0.005 
10 A  20 Hz - 45Hz  1 + 0.1  1.25 + 0.12  0.015 + 0.005 
  45 Hz - 2 kHz  0.35 + 0.1  0.5 + 0.12  0.015 + 0.005 
Notes: 