8808A 
Calibration Manual 
3-8 
Table 3-5. 2-Wire Ohms Verification Steps (cont.) 
90-Day Test Limit  1-Year Test Limit 
Nominal 
Input 
Range 
High Low High Low 
1.9 MΩ 2 MΩ STD+0.00063 MΩ
[1]
STD-0.00063 MΩ
[1]
STD+0.00084 MΩ
[1]
 STD-0.00084 MΩ
[1]
0 MΩ 20 MΩ 0.0006 MΩ -0.0006 MΩ 0.0006 MΩ -0.0006 MΩ 
19 MΩ 20 MΩ STD+0.0386 MΩ 
[1]
STD-0.0386 MΩ 
[1]
 STD+0.0481 MΩ 
[1]
 STD-0.0481 MΩ 
[1]
 
0 MΩ 100 MΩ 0.004 MΩ -0.004 MΩ 0.004 MΩ -0.004 MΩ 
100 MΩ 100 MΩ STD+1.504 MΩ 
[1]
 STD-1.504 MΩ 
[1]
 STD+1.754 MΩ 
[1]
 STD-1.754 MΩ 
[1]
 
[1]  5700 reading + resistance 
Direct Current Verification 
To verify the Amps dc function of the Meter, connect it to the test equipment as shown in 
either Figure 3-4 or Figure 3-5 depending on current level and apply the current levels 
listed in Table 3-6. 