Performance Verification and Calibration 
  Frequency Verification  3 
3-11 
Frequency Verification 
Use the function generator to apply the frequencies listed in Table 3-8. 
Table 3-8. Frequency Verification Steps 
Nominal Input  90-Day Test Limit  1-Year Test Limit 
Ampl. Freq. 
Range 
High Low High Low 
0.1 V  20 Hz  2000 Hz  20.04 Hz  19.96 Hz  20.06 Hz  19.94 Hz 
0.1 V  50 Hz  2000 Hz  50.01 Hz  50.00 Hz  50.01 Hz  50.00 Hz 
0.1 V  1900 Hz  2000 Hz  1900.19 Hz  1899.81 Hz  1900.19 Hz  1899.81 Hz 
0.1 V  19 kHz  20 kHz  19.0023 Hz  18.9977 Hz  19.0025 Hz  18.9975 Hz 
0.1 V  195 kHz  200 kHz  195.024 Hz  194.977 Hz  195.026 Hz  194.975 Hz 
0.1 V  1000 kHz  1000 kHz  1000.14 Hz  999.86 Hz  1000.16 Hz  999.84 Hz 
Adjustment (Calibration) 
Meter adjustments, or calibration, should be performed at the desired interval, or 
whenever a verification test indicates a Meter function is out of tolerance. The Meter 
accuracy will stay within specifications only if the adjustment procedure is performed at 
regular intervals. A one-year interval is adequate for most applications. The Meter’s 
accuracy specifications are not valid beyond the one-year interval. 
Equipment for Calibration 
The required equipment for calibration is the same as the equipment listed in Table 3-1. 
Adjustment Process 
The Meter adjustment process can either be managed through the front panel of the Meter 
or through the RS-232 port. In addition, a completely automated procedure is available as 
a Fluke Met/Cal program. 
Front Panel Adjustments 
To put the Meter in the calibration mode, remove the CAL seal and press the Cal Enable 
button located on the bottom-right side of the front panel with a thin needle. CAL will 
appear in the secondary display with six flashing dashes in the primary display. 
To calibrate a function: 
1.  Push A, B, D, E, F, or O to start the calibration process. A 
recommended calibration value will appear in the primary display and continue to 
flash until the input is set. 
Note 
The recommended value can be overwritten by pushing C for 
the + or – sign and a through f for the six digits. 
2.  With a stable input to the Meter, push R to start calibrating the current calibration 
point. The displayed value freezes and a flashing star appears in the display. The 
Meter automatically makes the necessary adjustment to bring the Meter into 
specification. No internal mechanical adjustments are necessary. 
The adjustment steps are divided into six areas by six functions. Table 3-9 lists the 
calibration steps and indicates the function/command, calibration point, required input 
signal with frequency, and a description.