Performance Verification and Calibration 
  Adjustment (Calibration)  3 
3-17 
Note 
When the Meter returns a recommended calibration value followed by 
“W4”, 4-wire compensation is enabled on the input. 
Note 
When in current calibration, if “Disconnect Test Leads!” is returned, all 
test leads should be removed from the Meter. 
Note 
When “Connect to mA socket!” is returned from the Meter, use the mA 
input terminal to connect to the calibrator. When “Connect to 10A socket” 
is returned from the Meter, use the 10A input terminal to connect to the 
calibrator. 
RS-232 Calibration (Automatic) 
Automatic calibration of the Meter is possible using a Fluke Met/Cal program. The 
remote program directs the calibration equipment to apply a series of shorts, opens, 
voltages, currents, and resistances to the input of the Meter. At each step, the Meter 
internally makes the necessary adjustment to bring the Meter into specification. No 
internal mechanical adjustments are necessary. 
Using an automated, computer-controlled procedure, the calibration and verification 
procedures can be performed on the Meter in less than 50 minutes. The 8808A 
MET/CAL procedure is available at no charge for MET/CAL Gold Support members 
from http://support.fluke.com/met-support-gold
. To purchase the procedure, contact 
Fluke’s MET/CAL Support group at metsupport@fluke.com
. 
Calibration Points 
Listed below are the calibration values required for each calibration point of the 
calibration procedure. 
  {//adc 
    {0.0,   -0.0001999,   -0.0001,   0.0001,   0.0001999},   range 1 
    {0.0,   -0.001999,    -0.001,    0.001,    0.001999},    range 2 
    {0.0,   -0.01999,     -0.01,     0.01,     0.01999},     range 3 
    {0.0,   -0.1999,      -0.1,      0.1,      0.1999},      range 4 
    {0.0,   -1.999,       -1.0,      1.0,      1.999},       range 5 
    {0.0,   -10.0,        -5.0,      5.0,      10.0},        range 6 
  }, 
  {//aac 
    {0.0,  0.001,   0.005,   0.01,    0.01999},             range 1 
    {0.0,  0.01,    0.05,    0.1,     0.1999},              range 2 
    {0.0,  0.01,    0.5,     1.0,     1.999},               range 3 
    {0.0,  1.0,     2.0,     5.0,     10.0},                range 4 
  },