Performance Verification and Calibration
Alternating Voltage Verification 3
3-5
Alternating Voltage Verification
To verify the Volts ac function of the Meter, connect it to the test equipment as shown in
Figure 3-1 and apply the voltage listed in Table 3-3.
Table 3-3. Alternating Voltage Verification Steps
Nominal Input 90-Day Test Limit 1-Year Test Limit
Ampl. Freq.
Range
High Low High Low
10 mV 50 Hz 200 mV 10.120 mV 9.890 mV 10.120 mV 9.880 mV
190 mV 20 Hz 200 mV 191.620 mV 188.380 mV 191.810 mV 188.190 mV
190 mV 45 Hz 200 mV 190.385 mV 189.615 mV 190.480 mV 189.520 mV
190 mV 20 kHz 200 mV 190.385 mV 189.615 mV 190.480 mV 189.520 mV
190 mV 50 kHz 200 mV 190.670 mV 189.330 mV 190.765 mV 189.235 mV
190 mV 100 kHz 200 mV 191.620 mV 188.380 mV 191.810 mV 188.190 mV
0.1 V 50 Hz 2 V 0.10115 V 0.09885 V 0.10120 V 0.09880 V
1.9 V 20 Hz 2 V 1.91620 V 1.88380 V 1.91810 V 1.88190 V
1.9 V 45 Hz 2 V 1.90385 V 1.89615 V 1.90480 V 1.89520 V
1.9 V 20 kHz 2 V 1.903985 V 1.89615 V 1.90480 V 1.89520 V
1.9 V 50 kHz 2 V 1.90670 V 1.89330 V 1.90765 V 1.89235 V
1.9 V 100 kHz 2 V 1.91620 V 1.88380 V 1.91810 V 1.88190 V
1 V 50 Hz 20 V 1.0115 V 0.9885 V 1.0120 V 0.9880 V
19 V 20 Hz 20 V 19.1620 V 18.8380 V 19.1810 V 18.8190 V
19 V 45 Hz 20 V 19.0385 V 18.9615 V 19.0480 V 18.9520 V
19 V 20 kHz 20 V 19.0385 V 18.9615 V 19.0480 V 18.9520 V
19 V 50 kHz 20 V 19.0670 V 18.9330 V 19.0765 V 18.9235 V
19 V 100 kHz 20 V 19.1620 V 18.8380 V 19.1810 V 18.8190 V
10 V 50 Hz 200 V 10.115 V 9.885 V 10.120 V 9.880 V
190 V 45 Hz 200 V 190.385 V 189.615 V 190.480 V 189.520 V
190 V 20 kHz 200 V 190.39 V 189.62 V 190.480 V 189.520 V
190 V 50 kHz 200 V 190.670 V 189.330 V 190.765 V 189.235 V
190 V 100 kHz 200 V 191.620 V 188.380 V 191.810 V 188.190 V
750 V 50 Hz 30 V 30.42 V 29.58 V 30.44 V 29.57 V
750 V 45 Hz 750 V 751.50 V 748.50 V 751.88 V 748.13 V
750 V 20 kHz 750 V 751.50 V 748.50 V 751.88 V 748.13 V
750 V 50 kHz 750 V 752.63 V 747.48 V 753.00 V 747.00 V
750 V 100 kHz 750 V 756.38 V 743.62 V 757.13 V 742.88 V