Fluke 8808A - Adjusting Meter Range and Rate

Fluke 8808A
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8808A
Calibration Manual
3-12
Table 3-9. Adjustment Steps
Step Func./Command Point Input Signal Description
Volts DC: zero points using the 4-wire short
1 VDC 0 mV/0V 0 mV Zero point of VDC, all ranges
Volts DC: Gains – Adjust using 5700A/5725A
2 VDC -199.9 mV -199.9 mV
3 VDC -100 mV -100 mV
4 VDC 100 mV 100 mV
5 VDC 199.9 mV 199.9 mV
Gain of VDC, 200 mV range
6 VDC -1.999 V -1.999 V
7 VDC -1 V -1 V
8 VDC 1 V 1 V
9 VDC 1.999 V 1.999 V
Gain of VDC, 2 V range
10 VDC -19.99 V -19.99 V
11 VDC -10 V -10 V
12 VDC 10 V 10 V
13 VDC 19.99 V 19.99 V
Gain of VDC, 20 V range
14 VDC -199.9 V -199.9 V
15 VDC -100 V -100 V
16 VDC 100 V 100 V
17 VDC 199.9 V 199.9 V
Gain of VDC, 200 V range
18 VDC -1000 V -1000 V
19 VDC -500 V -500 V
20 VDC 500 V 500 V
21 VDC 1000 V 1000 V
Gain of VDC, 1000 V range
Volts AC: Zero points – Adjust with 5700A in standby or disconnected
22 VAC 0 mV/0 V 0 mV Zero point of VAC, all ranges
Volts AC: Gains – Adjust using 5700A/5725A
23 VAC 10 mV 10 mV@1 kHz
24 VAC 100 mV 100 mV@1 kHz
25 VAC 150 mV 150 mV@1 kHz
26 VAC 199.9 mV 199.9 mV@1 kHz
Gain of VAC, 200 mV range
27 VAC 0.1 V 0.1 V@1 kHz
28 VAC 1 V 1 V@1 kHz
29 VAC 1.5 V 1.5 V@1 kHz
30 VAC 1.999 V 1.999 V@1 kHz
Gain of VAC, 2 V range