Performance Verification and Calibration
Adjustment (Calibration) 3
3-13
Table 3-9. Adjustment Steps (cont.)
Step Func./Command Point Input Signal Description
31 VAC 1 V 1 V@1 kHz
32 VAC 10 V 10 V@1 kHz
33 VAC 15 V 15 V@1 kHz
34 VAC 19.99 V 19.99 V@1 kHz
Gain of VAC, 20 V range
35 VAC 10 V 10 V@1 kHz
36 VAC 100 V 100 V@1 kHz
37 VAC 150 V 150 V@1 kHz
38 VAC 199.9 V 199.9 V@1 kHz
Gain of VAC, 200 V range
39 VAC 37.5 V 37.5 V@1 kHz
40 VAC 200 V 200 V@1 kHz
41 VAC 500 V 500 V@1 kHz
42 VAC 750 V 750 V@1 kHz
Gain of VAC, 750 V range
Frequency – Adjust using 5520A
43 FREQ 500 Hz 1.9 V@500 Hz
44 FREQ 1 kHz 1.9 V@1 kHz
45 FREQ 1.5 kHz 1.9 V@1.5 kHz
46 FREQ 5 kHz 1.9 V@5 kHz
47 FREQ 10 kHz 1.9 V@10 kHz
48 FREQ 15 kHz 1.9 V@15 kHz
49 FREQ 50 kHz 1.9 V@50 kHz
50 FREQ 100 kHz 1.9 V@100 kHz
51 FREQ 150 kHz 1.9 V@150 kHz
52 FREQ 250 kHz 1.9 V@250 kHz
53 FREQ 500 kHz 1.9 V@500 kHz
54 FREQ 750 kHz 1.9 V@750 kHz
Frequency
Ohms – Adjust using 5520A
55 OHMS
0 Ω 0 Ω, 4Wire Zero point of OHMS, 200 Ω
range
56 OHMS
50 Ω 50 Ω, 4Wire
57 OHMS
100 Ω 100 Ω, 4Wire
58 OHMS
150 Ω 150 Ω, 4Wire
Gain of OHMS, 200 Ω range
59 OHMS
1 Ω 1 Ω, 4Wire
60 OHMS
0.5 kΩ 0.5 kΩ, 4Wire
61 OHMS
1 kΩ 1 kΩ, 4Wire
62 OHMS
1.5 kΩ 1.5 kΩ, 4Wire
Gain of OHMS, 2 kΩ range