IDA-5
Users Manual
14
Table 8. Flow Screen Menu Controls
Channel 1 Flow
Average Flow 0.00
ml/h
Volume 0.00
Ml
Elapsed Time 00: 00: 00
Inst. Flow ml/h
Back Pressure mmHg
AutoStart Start End
Press ESC for Status Screen
gir09.eps
Menu Control Description
AutoStart
Starts the test when liquid movement is sensed. This is the preferred procedure of
operation.
Start
Starts the measurement immediately. This can find the start up qualities of an infusion
device.
End Cancels the test.
After you select AutoStart or Start and then push ENTER, the channel flow screen
changes to its active mode as shown in Table 9.
Table 9. Channel Flow Measurement Parameters
Channel 1 Flow
Average Flow 59.70
ml/h
Volume 19.60
ml
Elapsed Time 00: 20: 00
Inst. Flow 60.20 ml/h
Back Pressure 7 mmHg
Graph End
Press ESC for Status Screen
gir10.eps
Measurement Description
Average Flow Calculated flow from the volume delivered since the test started and the current time.
Volume Volume delivered since the test started.
Elapsed Time The time since the test started.
Inst. Flow
Calculated flow rate for the last measurement interval, or at slower flow rates the last
60 μl.
Back Pressure
Pressure at the inlet of the Product which is the pressure the infusion device pushes
against.