Infusion Device Analyzer
Utilities
27
The Product can automatically do tests in a specific manner when specified conditions
are met. Test-stop preferences specifies the default actions that are to occur. Default
values can be changed at the start of each test.
Set which step to edit from the drop-down list in the When field:
Volume When the set volume is reached. Zero disables this.
Time When the set time is reached. 00:00 disables this.
First Condition Whichever of volume or time is first reached.
Both Conditions When both the time and volume are reached.
The action when the condition is met is selected from the drop-down list presented by the
Reached then field:
No Action The test continues. The user is not told.
Notify Operator A beep sounds and a message shows.
Stop Test Only The test stops, a beep sounds, and a message shows.
Stop Test & Save The test stops and is saved. A message tells you the test
is saved.
The Preferred Parameters let you choose default pressure units and the position of
tolerance lines on flow graphs. Psi, kPa, and mmHg pressure units are available. The
percentage tolerance can be from 0 to 50 %.
Calibration
The Calibration menu gives access to calibration history and factory functions. Some
items on this menu are available to authorized-service personnel only. Each screen has
instructions on usage. “View Optics” is a diagnostic utility.
Edit Templates
See the “Templates” section earlier in this manual.