Infusion Device Analyzer
Specifications
31
Volume Measurement
Method .......................... Volume is measured directly by the measuring module
in minimum sample sizes of 60 μl
Range ............................. 0.06 ml to 9999 ml
Accuracy ....................... 1 % of reading ±1 LSD for flow rates of 16 to 200 ml/h
for volumes over 20 ml. otherwise 2 % of reading
±1 LSD for volumes over 10 ml under laboratory
conditions.
Max Test Duration ........ 100 hours
PCA Bolus/Dual Flow Measurement
Method ........................... See Volume measurement above
Min Bolus Volume ........ 0.5 ml
Resolution ...................... 60 μl increments
Max Test Duration ........ 100 hours
Pressure Measurement
Method (back Pressure
and Flow test) ................ Direct measurement of pressure at the inlet port.
Range ............................. 0 to 45 psi or equivalent in mmHg and kPa
Accuracy ........................ 1 % of Full Scale ±1 LSD under laboratory conditions
Max Test Duration ........ 1 hour
Other Specification
Templates ................................... Predetermined Test Sequences. Typical capacity 200
Storage of Results ....................... Test results stored for later viewing, printing or transfer
to PC. Typical capacity 250 tests.
Power down ................................ The results of tests in progress will be saved in
accidental power down
Computer Control ....................... The Product can be fully controlled using PC
Hydrograph and Ansur software for the Product
Applicable Standards
Safety ............................. IEC 61010 -1
EM Environment ........... IEC 61326-1: Portable
FCC CFR47 Part 15.107 and 15.109 Class A