Infusion Device Analyzer
Occlusion Tests
17
Occlusion Tests
When Occl is selected from the Device Information screen, the Occlusion Start Mode
screen in Table 12 is shown. Table 13 tells you about each field.
When the Occlusion Start Mode screen is first shown, Start briefly shows Wait as the
pressure circuit is zeroed.
Table 12. Occlusion Screen in Start Mode
Channel 1 Occlusion
Pressure 0.00
psi
Elapsed Time 00:00:00
Peak Pressure 0.00 psi
Time of Peak 00:00:00
Start End
Press ESC for Status Screen
gir13.eps
Menu Control Description
Start Starts the measurement.
End Stops the measurement.