MPS450 SERVICE MANUAL CALIBRATION
17
13) BLOOD PRESSURE 2 –– 5 µV 240 mm:
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 12 ± 0.012 mV.
¾ Select NEXT by pressing the [F4] key.
14) BLOOD PRESSURE 2 –– 40 µV Reference:
¾ Zero the DMM.
¾ Select NEXT by pressing the [F4] key.
15) BLOOD PRESSURE 2 –– 40 µV 240 mm:
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 96 ± 0.96 mV.
¾ Select NEXT by pressing the [F4] key.
16) BLOOD PRESSURE 3 –– 5 µV Reference:
¾ Move the MPS450 BP calibration cable to BP3. Keep the power supply and DMM
connections the same.
¾ Zero the DMM.
¾ Select NEXT by pressing the [F4] key.
17) BLOOD PRESSURE 3 –– 5 µV 120 mm:
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 6 ± 0.006 mV.
¾ Select NEXT by pressing the [F4] key.
18) BLOOD PRESSURE 3 –– 40 µV Reference:
¾ Zero the DMM.
¾ Select NEXT by pressing the [F4] key.
19) BLOOD PRESSURE 3 –– 40 µV 120 mm:
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 48 ± 0.048 mV.
¾ Select NEXT by pressing the [F4] key.