MPS450 SERVICE MANUAL CALIBRATION
23
V
ERIFICATION STEP CONNECTIONS DMM
SETTING
DISPLAY DMM
READING
BLOOD PRESSURE (Continued)
HIGH RANGE BASE Zero the DMM. BP3
40
µV/V/mmHg
Move the BP cal
cable to BP3.
300 mV DC
HIGH RANGE PEAK 46.8 – 49.2
LOW RANGE BASE Zero the DMM.
BP3
5
µV/V/mmHg
BP3
LOW RANGE PEAK 5.85 – 6.15
HIGH RANGE BASE Zero the DMM.
BP4
40
µV/V/mmHg
Move the BP cal
cable to BP4.
HIGH RANGE PEAK 39 – 41
LOW RANGE BASE Zero the DMM.
BP4
5
µV/V/mmHg
BP4
LOW RANGE PEAK 4.87 – 5.13
VERIFICATION OF CARDIAC-OUTPUT SPECS
SETUP
1. Connect the MPS450 temp/CO cable to CO/TEMP.
2. Connect the +DMM to the brown wire, and connect the –DMM to the blue wire.
STEPS
V
ERIFICATION STEP CONNECTIONS DMM
SETTING
DISPLAY DMM READING
CARDIAC OUTPUT
HIGH RANGE BASE 13.990 – 14.018 CO RESISTANCE (See SETUP, # 2.) 30 kΩ
HIGH RANGE PEAK 14.779 – 14.809