MPS450 SERVICE MANUAL CALIBRATION
20
VERIFICATION OF RESPIRATION-BASELINE SPECS
SETUP
1. Press the SETUP key.
2. Select RESP.
3. Select
BASE to select baseline impedance for the following steps.
STEPS
V
ERIFICATION STEP CONNECTIONS DMM
SETTING
D
ISPLAY DMM
READING
RESPIRATION-BASELINE IMPEDANCE
500 475 - 525
1000 950 - 1050
1500 1425 - 1575
RA-RL IMPEDANCE
RA-RL 3000 Ω
2000 1900 - 2100
500 475 - 525
1000 950 - 1050
1500 1425 - 1575
LA-RL IMPEDANCE LA-RL
2000 1900 - 2100
500 475 - 525
1000 950 - 1050
1500 1425 - 1575
LL-RL IMPEDANCE LL-RL
2000 1900 - 2100
V1-RL IMPEDANCE V1-RL
500 713 - 787
V2-RL IMPEDANCE V2-RL
500 713 - 787
V3-RL IMPEDANCE V3-RL
500 713 - 787
V4-RL IMPEDANCE V4-RL
500 713 - 787
V5-RL IMPEDANCE V5-RL
500 713 - 787
V6-RL IMPEDANCE V6-RL
500 713 - 787
VERIFICATION OF ECG-AMPLITUDE SPECS
SETUP
1. Press the SETUP key.
2. Select >.
3. Select UTIL.
4. Select DIAG.
5. Select PREV or NEXT until the LCD reads CAL TEST.