EasyManua.ls Logo

Fluke MPS450 - Troubleshooting Tips and Solutions

Fluke MPS450
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MPS450 SERVICE MANUAL TROUBLESHOOTING
26
TROUBLESHOOTING TIPS
FUNCTION: ECG
Symptom: The amplitude for Lead II is not correct.
Suggestion 1: The problem may be related to the monitor filters. Note that there will be a slight
loss of amplitude when simulating normal sinus rhythm. Use a pulse wave to verify
the amplitude more accurately.
Suggestion 2: The lead setting on the monitor may be incorrect. Check the monitor lead-select
switch to ensure that it is set to Lead II.
Suggestion 3: The leads may not be connected properly. Check the connections to the color-coded
ECG jacks.
FUNCTION: BLOOD PRESSURE
Symptom: There is no blood-pressure reading, or the reading is not as expected.
Suggestion 1:
The MPS450 transducer sensitivity may not be set correctly. Check the monitor
manufacturer’s BP-sensitivity requirements. Reset the MPS450 either to
5
µV/V/mmHg or to 40 µV/V/mmHg.
Suggestion 2: An incorrect blood-pressure cable may be in use. Note that many manufacturers use
plugs and/or configurations with wiring differences. Check the MPS450 BP-wiring
diagram, and compare it with the monitor specification.
Suggestion 3:
The monitor may not be zeroed. Zero the MPS450 BP channel. Then, zero the
monitor.
Suggestion 4: The monitor sensitivity (scaling) may not be correct. Adjust scaling on the monitor,
near the range simulated (e.g., 120 mmHg, etc.).
Suggestion 5: The performance waveform may be output currently, which turns off blood pressure.
Select the normal physiological waveform.
FUNCTION: RESPIRATION
Symptom: There is no respiration signal.
Suggestion 1: The correct respiration lead may not be selected. Check the respiration-lead
detection on the monitor, and switch accordingly on the MPS450.
Suggestion 2: The leads may not be connected properly. Verify that the color code is correct for
U.S. or International ECG leads.
Suggestion 3: The performance waveform may be output currently, which turns off respiration.
Select the normal physiological waveform.

Related product manuals