MPS450 SERVICE MANUAL CALIBRATION
22
RESPIRATION
HIGH RANGE BASE Zero the DMM. LEAD II
RESPIRATION
RA-LL 1 k
RESP HIGH PEAK 2.7 – 3.3
LOW RANGE BASE Zero the DMM.
LEAD I
RESPIRATION
LA-RA
RESP LOW PEAK 0.45 – 0.55
VERIFICATION OF BLOOD-PRESSURE SPECS
SETUP
1. Connect the MPS450 BP calibration cable to BP1.
2. Connect the +power supply to the brown wire, and the –power supply to the yellow wire.
3. Connect the +DMM to the brown wire, and connect the –DMM to the yellow wire.
4. Change the DMM setting to read 30 V DC with Null off.
5. Adjust the power supply (NOT the MPS450) for 10 ± 0.01 V.
6. Move the +DMM connection to the orange wire, and the –DMM connection to the blue wire.
7. Maintain the DMM and power-supply connections for all blood-pressure measurements.
NOTE: For some blood-pressure-specification tests, all that is required is to zero
the DMM, while other tests show a DMM reading and a tolerance.
STEPS
V
ERIFICATION STEP CONNECTIONS DMM
SETTING
D
ISPLAY DMM
READING
BLOOD PRESSURE
HIGH RANGE BASE Zero the DMM. BP1
40
µV/V/mmHg
BP1 300 mV DC
HIGH RANGE PEAK 156 – 164
LOW RANGE BASE Zero the DMM.
BP1
5
µV/V/mmHg
BP1
LOW RANGE PEAK 19.5 – 20.5
HIGH RANGE BASE Zero the DMM.
BP2
40
µV/V/mmHg
Move the BP cal
cable to BP2.
HIGH RANGE PEAK 92.8 – 98.8
LOW RANGE BASE Zero the DMM.
BP2
5
µV/V/mmHg
BP2
LOW RANGE PEAK 11.66 – 12.34