EasyManuals Logo

Fluke ScopeMeter 215C User Manual

Fluke ScopeMeter 215C
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
Calibration Adjustment
5.6 Final Calibration 5
5-13
5.6.5 Input A&B Volt Gain
Warning
Dangerous voltages will be present on the calibration source
and connection cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
Proceed as follows to do the Input A&B Volt Gain calibration.
1. Press
to select the first calibration step in Table 5-4.
2. Connect the test tool to the 5500A as shown in Figure 5-6.
al55ab.bmp
Figure 5-6. Test tool Input A&B to 5500 Normal Output
3. Set the 5500A to supply a DC voltage (NORMAL output), to the first calibration
point in Table 5-4.
4. Set the 5500A to operate (OPR).
5. Press
to start the calibration.
6. Wait until the display shows calibration status
:READY.
7. Press
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points of Table 5-4.
8. Set the 5500A to Standby, and continue at Section 5.6.6.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 215C and is the answer not in the manual?

Fluke ScopeMeter 215C Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 215C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals